<< Click to Display Table of Contents >> Navigation: DQM > DQM Checks > Radio Chemistry Estimated Results Check |
The DQM Check for Radio Chemistry Estimated Results checks for radio chemistry sample results between the minimum detectable concentration and the decision level concentration.
Check name is EarthSoft.DQM.RadChemEstimatedResults = Radiochemistry Estimated results.
This check uses three Global Parameters:
1.reportable_result – DT_RESULT.REPORTABLE_RESULT values to include in DQM datasets
2.result_type_code – DT_RESULT.RESULT_TYPE_CODE to include in DQM datasets
3.test_type_batch – AT_TEST_BATCH_ASSIGN.TEST_BATCH_TYPE to include in DQM dataset.
There are two rules:
1.MDC < RESULT < 3*MDC.
2.MDC < RESULT < 3*TPU(1 sigma).
Qualify a result with the first rule if, in DT_RESULT:
•MINIMUM_DETECTABLE_CONC < RESULT_NUMERIC < 3 x MINIMUM_DETECTABLE_CONC.
Qualify a result with the second rule if, in DT_RESULT:
•MINIMUM_DETECTABLE_CONC < RESULT_NUMERIC < 3 x TPU(1 sigma),
where TPU(1 sigma) = RESULT_ERROR_DELTA if UNCERTAINTY = "1 sigma", or
where TPU(1 sigma) = RESULT_ERROR_DELTA / 1.96 if UNCERTAINTY = "2 sigma".
In the Errors tab of the DQM Event Review Form these errors may be listed:
1."One of the required rules for this check is not present in the RT_DQM_RULE table."
If you receive this error, the QAPP can be updated following the How to Update Existing QAPPs documentation page.
2."{FIELDNAME} is NOT Numeric! So, Estimated Result check cannot be completed for Sample, Method, CAS_RN"
3."The UNCERTAINTY value of {0} is neither 1 sigma nor 2 sigma."
If you receive this error, the check will be skipped.
This check will only qualify records where MINIMUM_DETECTABLE_CONC, RESULT_ERROR_DELTA, and UNCERTAINTY are not NULL.
Copyright © 2023 EarthSoft, Inc • Modified: 24 May 2022